Model CFP- 15M
Convenience and Independence
The Model CFP-15M combines the high pressure and
variable flow rate available from an oxygen cylinder with
the oxygen generation and replenishment of an oxygen
concentrator. Empty oxygen cylinders become a thing of
the past. By simply connecting an empty oxygen cylinder(s)
(up to two H-size at once) to its discharge line and turning
it on, it will automatically re-fill them to 2200 psi.
The Model CFP-15M turns itself off once the desired level
of pressure is reached. It will top off most cylinders in a few hours, providing independence from traditional oxygen
supply channels.
These mobile oxygen generation systems have the added benefit of being easily transported to battlefield hospitals. Unlike other portable systems, the Model CFP-15M can be assembled onto its own cart for mobility inside a hospital tent. This feature saves space, and eliminates the tripping hazard of other portable systems for (2) K-size cylinders, and a station for transfilling to smaller oxygen cylinders. Field assembly and set-up of the Model CFP-15M can be done in minutes.
Oxygen Quality
The Oxygen produced by this system will be of Medical Grade, conforming to the United States Pharmacopeia Standard for ‘Oxygen 93 Percent’. There is an oxygen purity monitor which operates continuously. Air and oxygen pressure is also monitored.
Reliability
A password secured touch-screen monitor provides easy access to all system controls and readouts. The system is designed to run automatically and requires little attention once it is started. It has an automatic shutdown feature if oxygen purity drops, with display and audible alarms. The CFP-15M incorporates pressure swing adsorption (PSA) technology for oxygen generation. PSA oxygen generators have been used in many critical applications for years.
The design is modular, enabling quick and easy service. Should it be necessary, any of the major subsystems can be changed within minutes.
Model CFP-15M Specifications
| Product Gas: | | Oxygen Quality | Medical Grade - Conforming to USP for
'Oxygen 93 Percent' | | Oxygen Concentration |
93%(± 3%) | | Carbon Dioxide Limit |
³300 ppm | | Carbon Monoxide Limit |
³10 ppm | | Oxygen Dewpoint | -60°F (-51°C) | | Oxygen Flow Rate |
0-15 SCFH
(0-7 LPM) | | Oxygen Pressure |
up to 2200 psi (152 bar) |
*A standard cubic foot is a cubic foot of gas at 14.7 psia (sea level atmospheric pressure) and 60 Degrees Farenheight.
System Operation: | Feed Air Flow Rate | Standard Unit Includes Feed Air Systems
4 SCFM @ 40 psi
| Feed Air Quality | ISO 8573.1 Class 4 Minimum or Grade D (see below) | Sound Level | Door Open: 68 dBA @ 1 m
Door Closed: 62 dBA @ 1 m | Power Supply | 115 VAC, 10 A, 60 Hz - standard |
Measurements and Weights: | | OGS Unit (top) | 21" x 20" x 33" (W x D x H)
53 cm x 50 cm x 86 cm (W x D x H)
88 lb (40kg) | OCS Unit (bottom) |
19" x 20" x 23" (W x D x H)
47 cm x 51 cm x 58 cm
118 lb (54 kg) | | Cart Only |
27" x 36" x 39" (W x D x H)
68 cm x 92 cm x 100 cm
(W x D x H)
96 lb (44 kg) | Cylinder Connection | CGA 540 Standard/CGA 870 Optional | ISO 8573.1 Air Quality Specifications: | Maximum Partivle Size | 15 Microns | Dewpoint | 37OF (2.8OC) (940 ppmv) | Cylinder Connection | 5 ppmv |
|